Gas Supply:
Gas Source: O2,or O2&Air(Option), or O2&N2O(Option);
Gas Source Pressure Gauge: O2,or O2&Air(Option), or O2&N2O(Option); Pressure of Pipeline Inlet:
O2,280kPa~600kPa/41Psi~87Psi, less than 180L/min;N2O(Option),280kPa~600kPa/41Psi~87Psi,less than180L/min; Air(Option),280kPa~600kPa/41Psi~87Psi, less than 180L/min;Security Valve: release pressure less than 10 kpa;O2 Lost Alarm(Option):Sound alarm, alimumun cylinder with whistle;N2O cut off valve protection(Option): when O2 is lost, N2O will be cut off.
Flow Meter:
2 tubes high precision flow meter,0~1L/min,1.1L/min~10L/min; or 0L/min~10L/min(O2&Air or O2&N2O);Precision: At 20 degree, atmospheric pressure is 101.3kPa, the precision of flow meter is level:2.5;O2&N2O Linkage Device(O2&N2O,Option):make insure O2 concentration is not lower than 25%;
O2 Flush:35L/min~75L/min;
One Key Switch: Switch between rebreath and Non-Rebreath;
ACGO Port: Non-Rebreath, with an alone airway pressure gauge;
Selectatec Mounting Bar:single place used for vaporizer, it is easy to replace;
Vaporizer:Anesthesia Vaporizer: one of Isoflurane/Halothane/Sevoflurane;
Respiration System/CO2 Circle Absorber/Rebreath:
Soda LIme Chamber: 800mL;
Pop Off Vlave: with one key to close; APL valve(Option): 0.19~6kPa;Airway Pressure Gauge:-20-100cmH2O;
Dimension&Weight:
Table Top:Width*Deepth*Height:45cm*25cm*40cm,22kg;With Trolley Stand:60cm*60cm*130cm,28kg;
Standard Configuration:
Main unit, one unit X60 Isoflurane Anesthesia Vaporizer, one unit O2 gas source pipeline, 0.5,1,2,3L airbag, one set big breath circuit, one set small breath circuit.
Package:
Main unit with one anesthesia vaporizer with necessary accessories:
Lengh*Width*Height=95cm*30cm*55cm, Gross weight:26kg;
Trolley(Option): Lengh*Width*Height=70cm*57cm*26cm,Gross weight:10kg.